Biomedical Science Letters

eISSN 2288-7415

Table. 1.

Comparison of analytical performances between commercial and MD well plate

Performance data Third party well plate MD well plate
LOD-LOB+1.645 (blank SD) No. 1, 2, 3 average LOD = 91 pg/mL No. 1, 2, 3 average LOD = 37.6 pg/mL
Analytical sensitivity 91 pg/mL 38 pg/mL
Measuring range 100~12,000 pg/mL 40~12,000 pg/mL
Reaction volume 100 μL 50 μL
Reaction time (based on 4 levels) 30 min 30 min

Reproducibility Inter-assay CV Intra-assay CV Inter-assay CV Intra-assay CV

< 15% < 10% < 11% < 9%
Biomed Sci Letters 2024;30:131-6 https://doi.org/10.15616/BSL.2024.30.3.131
© 2024 Biomed Sci Letters